In Vivo Evaluation of Customized Pulse Oximeter and Sensitivity Pulp Tests for Assessment Of Pulp Vitality.

In Vivo Evaluation of Customized Pulse Oximeter and Sensitivity Pulp Tests for Assessment Of Pulp Vitality.

J Clin Pediatr Dent. 2018 Dec 06;:

Authors: Sharma DS, Mishra S, Banda NR, Vaswani S

Abstract
OBJECTIVE: The purpose of this study was to test a customized pulse oximeter (CPO) for evaluation of pulp vitality in primary and permanent teeth against clinical diagnosis (vital and untreated non-vital) in order to expand its clinical use for pulp preservation.
STUDY DESIGN: CPO was evaluated on intact primary and permanent central or lateral incisor (CI, LI) teeth- vital (group 1, 20n each); untreated non-vital (group 2, 10n each) and; root filled non-vital (group 3, 10n each) of children 4-12 years according to inclusion/exclusion criteria. For each patient CPO was first applied on finger followed by vitality tests in following sequence-electrical, pulse oximetry and thermal tests.
RESULTS: Mean oxygen saturation (%SpO2) in permanent and primary-vital teeth was 88.78% & 87.77% respectively; non-vital teeth was 74.67% & 75.00% respectively; and in all root filled teeth was 0%. Tooth and finger oxygen saturation values showed strong positive relationship in vital primary or permanent teeth and; no correlation in untreated non-vital primary or permanent teeth. The accuracy rate of thermal pulp test and pulse oximetry was 100% and for electrical pulp test it was 90% for permanent and 86.67% for primary teeth.
CONCLUSION: The CPO tested in this study proved to be a valuable adjunct for diagnosing pulp vitality by objective means.

PMID: 30520699 [PubMed - as supplied by publisher]