Blog Archive

Αλέξανδρος Γ. Σφακιανάκης

Sunday, November 29, 2020

Quantification of DNA through the NanoDrop Spectrophotometer: Methodological Validation Using Standard Reference Material and Sprague Dawley Rat and Human DNA

alkiviadis.1961 shared this article with you from Inoreader

ijac.banner.jpg

This study aimed to validate an analytical method to determine DNA concentration using standard reference material (NIST SRM 2372) and Sprague Dawley rat and human DNA. Microvolumes were used to analyse DNA samples. Linearity showed correlation coefficients higher than R ≥ 0.9950, and the precision value was ≤2% CV. Trueness based on bias and the percentage of recovery showed bias values lower than Z-test with a 95% confidence level and a recovery percentage within the range (% Rec = 100% ± 5%), and the stability of the samples was 60 days (2–4°C).
View on the web

No comments:

Post a Comment